Measured values | O2 |
Measurement principles | Diode laser spectroscopy (TDLS) |
Measuring ranges | | O2 | 0 ... 5 Vol.-% / 0 ... 100 Vol.-% |
|
Response time (t90) | ≤ 10 s |
|
Zero point drift | | | ± 0.1 Vol.-%: per year |
|
Process temperature | | | –20 °C ... +80 °C |
|
Process pressure | | | 800 hPa ... 1,400 hPa |
|
Ambient temperature | | | –20 °C ... +60 °C |
|
Storage temperature | | | –30 °C ... +80 °C |
|
Ex-approvals | | IECEx | Sender/receiver unit: Ex op is / ib IIB T4 Ga/Gb; Ex ib tb op is IIIC T85°C Db | | TSA151 connecting unit: Ex eb mb [ib] IIB T4 Gb; Ex tb [ib] IIIC T85°C Db | ATEX | Sender/receiver unit: II 1/2 G Ex op is / ib IIB T4 Ga/Gb; II 2 D Ex ib tb op is IIIC T85°C Db | | TSA151 connecting unit: II 2 G Ex eb mb [ib] IIB T4 Gb; II 2 D Ex tb [ib] IIIC T85°C Db | NEC/CEC (US/CA) | Sender/receiver unit: Class I, Division 2, Group A, B, C and D; T4 Ta=-20°C to +60°C; IP66 | | Measuring probe: Class I, Division 1, Group A, B, C and D; T4 Ta= –20°C to +80°C; IP66 |
|
Electrical safety | CE, FM |
|
Analog outputs | 1 output active: 0/4 ... 20 mA, 500 Ω 1 output active: 0/4 ... 20 mA, 200 Ω Only for ATEX/IECEx version TRANSIC151LP |
Digital outputs | 1 relay contact: 30 V AC, 1 A / 60 V DC, 0.5 A 1 NAMUR output: Only for ATEX/IECEx version TRANSIC151LP |
Serial | ✔ , ✔ | Remark | Not for the ATEX/IECEx version | Type of fieldbus integration | RS-232 | | RS-485 | Function | Proprietary service interface |
|
USB | ✔ | Remark | Not approved for Ex-applications | Function | Proprietary service interface |
|
Dimensions (W x H x D) | | | Dimensions may vary. For details, see the dimensional drawings. |
|
|
Material in contact with media | Stainless steel 1.4404 (AISI 316L), Kalrez®, EPDM, PTFE, SiN, MgF2, polymer coating |
Mounting | Flange for direct installation in the process Wall-mounting bracket for ambient air measurement Wall-mounting bracket for measurement with measuring gas cell |
Energy supply | | Voltage | 24 V DC, ± 10 % | | TRANSIC151LP 21.6 V DC ... 26.4 V DC, ± 10 % | | For ATEX/IECEx version via TSA151 connecting unit, a PELV power supply is mandatory | Current consumption | ≤ 500 mA | Power consumption | ≤ 6 W | | TRANSIC151LP: ≤ 5.2 W |
|
Corrective functions | Adjustment with ambient air or test gases |
Test functions | Contamination check |