Measured values | CO, NO, O2 |
Performance-tested measurands | CO, NO, O2 |
Maximum number of measurands | 3 |
Measurement principles | NDIR spectroscopy, Paramagnetic dumbbell principle |
|
Measuring ranges | | CO | 0 ... 50 ppm | NO | 0 ... 1,500 ppm | O2 | 0 ... 25 Vol.-% |
|
Certified measuring ranges | | CO | 0 ... 2,500 mg/m³ | O2 | 0 ... 25 Vol.-% |
|
Response time (t90) | ≤ 2 s |
Sensitivity drift | | | 1 %: of measuring range full scale |
|
Process temperature | | | ≤ +1,400 °C |
|
Process pressure | | | –100 hPa ... 100 hPa Relative |
|
Process gas humidity | | | ≤ 90 % |
|
|
Ambient temperature | | | 0 °C ... +40 °C |
|
Storage temperature | | | –20 °C ... +70 °C |
|
Ambient pressure | | | 900 hPa ... 1,100 hPa |
|
Ambient humidity | | | ≤ 85 % Relative humidity |
|
|
Analog outputs | 3 outputs: 4 ... 20 mA |
Analog inputs | Optional 2 inputs possible |
Digital outputs | Optional 4 outputs possible |
Digital inputs | Optional 4 inputs possible |
Dimensions (W x H x D) | | Analyzer cabinet | 1,200 mm x 2,000 mm x 800 mm | Heat exchanger cabinet | 1,100 mm x 2,200 mm x 800 mm | Gas sampling unit | 1,500 mm x 1,000 mm x 4,000 mm |
|
Material in contact with media | PVDF, stainless steel 1.4404 (AISI 316L) |
Material | Steel sheet enclosure |
Energy supply | | Three-phase current | 3-phase: 400 V, 50 Hz | | 3-phase: 400 V, 60 Hz | Voltage | 380 V AC | Frequency | 50 Hz |
|
Auxiliary gas connections | | Instrument air | 1/4″ |
|
Corrective functions | Manual or automated adjustment with test gases |
System components | Analyzer cabinet Heat exchanger cabinet SMB-9004H-01 sampling probe Analyzer |