alt
alt

Sick-System MCS300P HW
System MCS300P HW

Gas AnalyzersMCS300P HW, System MCS300P HW

Gas Analyzers
MCS300P HW, System MCS300P HW

Measured valuesCO, CO2, HCl, H2O, NH3, NO, NO2, N2O, SO2, O2
Measurement principlesInterference filter correlation, Gas filter correlation, Zirconium dioxide sensor
Gas flow rate
150 l/h ... 800 l/h
Measuring ranges
CO0 ... 800 ppm / 0 ... 5 Vol.-%
CO20 ... 20 Vol.-% / 0 ... 100 Vol.-%
HCl0 ... 700 ppm / 0 ... 12,000 ppm
H2O0 ... 1 Vol.-% / 0 ... 60 Vol.-%
NH30 ... 550 ppm / 0 ... 5,000 ppm
NO0 ... 750 ppm / 0 ... 15,000 ppm
NO20 ... 500 ppm / 0 ... 4,000 ppm
N2O0 ... 500 ppm / 0 ... 7,500 ppm
O20 ... 1 Vol.-% / 0 ... 21 Vol.-%
SO20 ... 200 ppm / 0 ... 10 Vol.-%
Other measuring ranges and components on request
2 measuring ranges per component
Response time (t90)≤ 150 s
Depending on type and length of sample gas line
Zero point drift
IR< 2 %: of measuring range full scale per week
Detection limit
≤ 2 %: of measuring range full scale
Process temperature
≤ +1,300 °C
Depending on sampling probe
Sample gas temperature
≤ +220 °C
Process pressure
800 hPa ... 1,200 hPa
Ambient temperature
+5 °C ... +35 °C
Temperature change maximum ±10 °C/h
With cooling device+5 °C ... +45 °C
Temperature change ±10 °C/h maximum; other temperatures on request
Ambient humidity
≤ 80 %
Non-condensing
Electrical safetyCE
Enclosure rating
IP43
Analog outputs2 outputs:
0/4 ... 22 mA, 500 Ω
Electrically isolated; additional outputs if using I/O modules (option)
Analog inputs2 inputs:
0/4 ... 22 mA, 100 Ω
Electrically isolated; additional inputs if using I/O modules (option)
Digital outputs

3 potential-free contacts:
Additional outputs if using I/O modules

2 power relays:
Electrically isolated; additional outputs if using I/O modules (option)

Digital inputs2 potential-free contacts:
Additional inputs if using I/O modules
Modbus
Type of fieldbus integrationTCP
Ethernet
FunctionConnection to SOPAS ET software or OPC server
Indication
LC display
Status LEDs: “Power,” “Failure,” and “Maintenance request”
InputFunctional keys
Operation

Via control unit in the analyzer

Two operating levels, one password-protected

Application specific programs for system control

Via software SOPAS ET

ModelSteel sheet cabinet
Dimensions (W x H x D)
800 mm x 2,160 mm x 600 mm (for details see dimensional drawings)
Weight
≤ 300 kg
Energy supply
Depending on device version
Corrective functions

Cross-sensitivity compensation of up to 6 interferents

Internal adjustment unit (option)

Test functionsAutomatic check cycle for zero and span point

As Mnelko, we are a reliable partner in industrial automation solutions, offering high-quality and innovative products with a wide product range to meet all the needs of your business. Our product portfolio consists of solutions designed based on the principles of high performance, durability, and efficiency to support modern industrial processes.
All of our products ensure maximum reliability even in challenging working conditions and help your business operations run smoothly. The products we offer in industrial automation, energy management, cabling solutions, and many other areas adapt flexibly to the needs of different sectors.
Additionally, our products are manufactured using only high-quality materials and comply with international standards. Through the solutions we provide, we enable our customers to increase operational efficiency and optimize costs. Our company closely follows technological advancements and continuously offers innovative products to help our customers gain a competitive advantage.
On each of our product pages, you can find comprehensive information about technical details, areas of use, and product features. You can explore all the products you need to strengthen your industrial processes on our website and enjoy a seamless purchasing experience.

Similar Products

Contrinex-YXL-0001-000
Contrinex
Contrinex-IDLR-C23PP-NMS-A0
Contrinex
Contrinex-IDLR-M18MP-NMS-A0
Contrinex
Sick-PBS-CB016SN1SS0BMA0Z
Sick
Sick-PBS-RB6X0SGESS0CMA0Z
Sick
Sick-PBS-CK4X0SG1SSNGMA0Z
Sick
Sick-PBS-RB040SF2FSNNMA0Z
Sick
Sick-PBS-CB1X0SG2SS0BMA0Z
Sick
Sick-PBS-RB4X0SGHSS0AMA0Z
Sick
Sick-PBS-RB400SG2SS0CMA0Z
Sick
Sick-SRM50-HGV0-S23
Sick
Sick-PBS-RB4X0SG1SSNE5A0Z
Sick