| Measured values | CH4, C2H6, C3H8, Corg, CO, CO2, HCl, HF, H2O, NH3, NO, NO2, N2O, O2, SO2, gas velocity, Temperature, pressure, dust concentration |
| Performance-tested measurands | CO, CO2, HCl, H2O, NO, NO2, O2, SO2, gas velocity, Temperature, pressure, dust concentration |
| Maximum number of measurands | 13 |
| Measurement principles | FTIR spectroscopy |
| Length of measuring path | | | 8.48 m |
|
|
| Measuring ranges | | | CH4 | 0 ... 50 mg/m³ / 0 ... 150 mg/m³ | | SO2 | 0 ... 75 mg/m³ / 0 ... 1,500 mg/m³ | | C3H8 | 0 ... 55 mg/m³ | | Corg | 0 ... 15 mg/m³ / 0 ... 500 mg/m³ | | CO | 0 ... 75 mg/m³ / 0 ... 1,500 mg/m³ | | CO2 | 0 ... 25 Vol.-% | | HCl | 0 ... 15 mg/m³ / 0 ... 150 mg/m³ | | H2O | 0 ... 40 Vol.-% | | NH3 | 0 ... 10 mg/m³ / 0 ... 50 mg/m³ | | NO | 0 ... 200 mg/m³ / 0 ... 2,000 mg/m³ | | NO2 | 0 ... 100 mg/m³ / 0 ... 500 mg/m³ | | N2O | 0 ... 50 mg/m³ / 0 ... 500 mg/m³ | | O2 | 0 ... 21 Vol.-% |
|
| Certified measuring ranges | | | H2O | 0 ... 40 Vol.-% | | NO | 0 ... 500 mg/m³ | | O2 | 0 ... 25 Vol.-% | | SO2 | 0 ... 600 mg/m³ | | Dust concentration | 0 ... 100 mg/m³ | | Gas velocity | 0 ... 40 m/s |
|
| Response time (t90) | ≤ 200 s |
| Accuracy | | | ≤ 2 % Of measuring range full scale |
|
| Sensitivity drift | | | < 3 %: of the measuring range full scale value per maintenance interval |
|
| Zero point drift | | | < 0.5 %: of the measuring range full scale value per maintenance interval |
|
| Reference point drift | | | ≤ ± 2 %: of the measuring range full scale value per month |
|
| Detection limit | | | < 2 %: of measuring range full scale |
|
| Process temperature | | | ≤ +1,300 °C |
|
| Sample gas temperature | | | ≤ +220 °C |
|
| Process pressure | | | 900 hPa ... 1,100 hPa |
|
| Process gas humidity | | | Non-condensing |
|
| Ambient temperature | | | +5 °C ... +35 °C |
|
| Storage temperature | | | –20 °C ... +60 °C |
|
| Ambient pressure | | | 900 hPa ... 1,100 hPa |
|
| Ambient humidity | | | ≤ 80 % Relative humidity; non-condensing |
|
| Conformities | HJ/T 76-2007 |
|
| Analog outputs | 4 outputs: 0/4 ... 20 mA, + 500 Ω |
| Analog inputs | 4 inputs: 0/4 ... 20 mA |
| Digital outputs | 64 outputs: + 48 V AC, 0.5 A, 35 W / 48 V DC, 0.5 A, 24 W |
| Digital inputs | 64 inputs: + 3.9 V, 4.5 mA, 0.55 W |
| Additional inputs | For temperature, pressure, gas velocity, and dust concentration |
|
| Input | Touchscreen |
| Operation | Menu-driven operation via control unit |
| Menu languages | English |
| Model | Indoor analyzer cabinet |
| Dimensions (W x H x D) | | | 800 mm x 2,200 mm x 800 mm |
|
|
| Material in contact with media | PTFE, Calcium fluoride |
| Energy supply | | | Voltage | 220 V AC | | Frequency | 50 Hz | | Current consumption | Depending on type and length of sample gas line | | Power consumption | Analyzer: ≤ 1,000 VA | | Sampling probe: ≤ 450 VA |
|
| Sample gas connections | | | Swagelok 8 mm |
|
| Auxiliary gas connections | | | DN 6/8 |
|
| Corrective functions | Manual or automated adjustment with test gases |
| Test functions | Internal zero and reference point check |
| System components | Analyzer Sampling probe Sample gas line Control unit |
| Integrated components | Ejector pump |
| Options | Zirconium dioxide sensor (oxygen measurement) |